Productinformation

Hanna monitor for marine aquariums pH, salinity, temperature (HI981520)

Product no.: 141471

The system can be calibrated at one or two points for pH and at one point for conductivity. Seawater salinity is expressed in either parts per thousand (ppt), Practical Salinity Units (PSU), or Specific Gravity (S.G.). A high/low alarm can be configured for each parameter and the results are displayed on a large, easy-to-read LCD.

The attached two-electrode conductivity probe features a built-in temperature sensor for rapid, high-accuracy temperature readings.

The Hanna HI1286 pH double junction, gel-filled pH electrode features a PEI resin body that is easy to clean and resistant to many aggressive chemicals. Additionally, the double junction design is less susceptible to clogging for fast and stable readings.

Two Alarm operating modes (High/Low Value) for each supported parameter

  • High/Low Alarm generated when measured value exceeds or drops below configured parameter High/Low Value
  • Acoustic signal generated each time an alarm is triggered
  • All readings are compensated for variations in temperature
  • Temperature is displayed in °C or °F along with pH and EC readings
  • The EC probe contains an integral temperature sensor that simplifies installation
  • Easy to read LCD display

Specifications

pH
Range*0.0 to 14.0 pH
Resolution0.1 pH
Accuracy±0.2 pH
CalibrationManual or automatic two-point calibration in 7.01 and 10.01 pH
ppt
Range0.0 to 70.0 ppt (g/L
Resolution0.1 ppt (g/L)
Accuracy±1.0 ppt between 0.0 ppt and 40.0 ppt
Accuracy±2.0 ppt between 40.0 ppt and 70.0 ppt
CalibrationAutomatic, single point in 35.00 ppt
PSU
Range0.0 to 70.0 PSU
Resolution0.1 PSU
Accuracy±1.0 PSU between 0.0 PSU and 40.0 PSU
Accuracy±2.0 PSU between 40.0 PSU and 70.0 PSU
Temperature
Range0.0 to 50.0 °C (32.0 to 122.0 °F)
Resolution0.1 °C / 0.1 °F
Accuracy±0.5 °C / ±1.0 °F
CalibrationAutomatic, 5 to 50 °C (41 to 122 °F)
Alarm
pHHigh or Low with Enable or Disable option
ECHigh or Low with Enable or Disable option
TemperatureHigh or Low with Enable or Disable option
with High/Low alarmEnabled
pH ProbeHI1286
ECAttached
Power supply12 Vdc adapter (included) from 115 Vac, and 230 Vac
Environment0 to 50°C (32 to 122°F); RH max 95%, non-condensing
CasingP65 ingress protection
Dimensions125 x 185 x 38 mm (4.92 x 7.28 x 1.49”)
Weight300 g (10.6 oz)
* The range may be limited by the probe’s limits